VICTOR 05 Loop Calibrator
Input
|
Range
|
Input range
|
Resolution
|
Accuracy
|
Explanation
|
Voltage
|
28V
|
-0.200 ~28.000V
|
1mV
|
±0.02% reading ±2mV
|
Input resistance 2MΩ
|
Current
|
20mA
|
-1.000~22.000mA
|
0.001mA
|
±0.02% reading ±4uA
|
Input resistance 20MΩ
|
Loop current
|
20mA
|
0.000~22.000mA
|
0.001mA
|
±0.02% reading ±4uA
|
Supply 24V loop power
|
Output
|
Range
|
Input range
|
Resolution
|
Accuracy
|
Explanation
|
Current
|
20mA
|
0.000~22.000mA
|
0.001mA
|
±0.05% Setting value ±4uA
|
20mA,the max.overload is 1kΩ(Note1)
|
Analogue transducer
|
-20mA
|
0.000~-22.000mA
|
0.001mA
|
±0.05% Setting value ±4uA
|
20mA,the max.overload is 1kΩ(Note2)
|
Loop power
|
24V
|
|
|
±10%
|
The max.output current is 25mA
|
Power
|
9V battery (ANSI/NEDA 1640A or IEC 6LR619 alkaline) or AC power adapter
|
Battery
|
Approx. 20 hours under the condition of 10mA
|
Max.allowable voltage
|
30V
|
Operation temperture range
|
0-50°C
|
Operation humidity range
|
≤80%RH
|
Store temperature range
|
≤-10°C-50°C
|
Store humidity range
|
≤90%RH
|
Size
|
200×100×40mm
|
Weight
|
550g(with holster)
|
Accessories
|
Operation manual,test line CF-36(probe with alligator clip)
|
Options
|
AC adapter(VCPS),test line CF-31-A(probe clip)
|
Security
|
Conforms to IEC 1010
|
Note:
1.The max.overlaod is 1kΩ at 20mA range when the power is higher than 6.8V,
The max.overlaod is 700Ω at 20mA range when the power is higher than 5.8-6.8V,
2.Power supply range:DC 5-25V
3.Temperature coefficient:±0.005%,range/°C(5°C-18°C,28°C-40°C)